BioSigRP freezes or displays a flat line in the multi-plot during averaging
Tech Note: TN0126
Product: BioSigRP
Version: All
Date Added: 2004-05-04
Issue
BioSigRP might freeze or display a flat line in the Running Average or EEG plot during averaging, if:
-
The preamp is not connected and turned on.
-
The stimulus or acquisition circuit (RCO file) is not compatible with the selected device.
-
The configuration file is damaged or corrupted.
-
Channel 2 circuit components are erroneously being read and preventing the SGI from completing.
-
Length of sound output sig file for Channel 1 does not match length of Channel 2 file.
Solution
The stimulus or acquisition circuit (RCO file) is not compatible with the selected device:
Problems will occur if the sampling rate and/or the number of analog I/O ports accessed in the RCO does not match the acceptable sampling rate or number of analog I/O ports for the device. For example, the Record25KX4.rco includes four channels of acquisition and the RP2.1 has only two analog inputs. Also, the Record50KX2.rco uses a sampling rate of 50K and the RA16 accepts a maximum sample rate of 25K on the A/D inputs.
The configuration file is damaged or corrupted:
The configuration file can be damaged or corrupted anytime BioSigRP crashes. Using a damaged file can cause a variety of problems and might cause the software to crash. Recreating and using a new, but identical, configuration file can eliminate the problem. If this solves your problem be sure to delete the corrupted or damaged configuration file.